29
Cyanuric Acid
CYANURIC ACID
SPECIFICATIONS
Range 0 to 200 mg/L
Resolution 1 mg/L from 0 to 100 mg/L;
10 mg/L above 100 mg/L
Accuracy ±1 mg/L ±15% of reading at 25 °C
Typical EMC ±1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the turbidimetric method. The reaction between cyanuric acid and the
reagent causes a white suspension in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93722-0 Powder reagent 1 packet
REAGENT SETS
HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 34.
MEASUREMENT PROCEDURE
Select the
Cyanuric Acid
method using the procedure described in the
Method Selection
section (see page 15).
Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL