Página 1 - HI 83213

1Instruction ManualHI 83213Multiparameter BenchPhotometerfor Municipal Wastewaterwww.hannainst.com

Página 2 - TABLE OF CONTENTS

10• Pay attention to push the cuvette completely down in the holder and to align the white point on thecap to the indexing mark on the meter.• In orde

Página 3 - GENERAL DESCRIPTION

11METHOD REFERENCE TABLEMethod Method Pagedescription1 Aluminum 172 Ammonia MR 193 Ammonia LR 21 4 Bromine 235 Free Chlorine 256 Total Chlorine 2

Página 4 - PRECISION AND ACCURACY

12POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note: Alw

Página 5

13DATA MANAGEMENTThe instrument features a data log function to help you keep track of all your analysis. The data log canhold 200 individual measurem

Página 6 - FUNCTIONAL DESCRIPTION

14SETUPSETUPSETUPSETUPSETUPIn the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuringsequence and others are g

Página 7 - KEYPAD DESCRIPTION

15Date / TimeThis option is used to set the instrument’s date and time.Press the Modify key to change the date/time.Press the W X functional keys to h

Página 8

16HI 83213 offers an interactive contextual help mode that assists the user at any time.To access the help screens press HELP.The instrument will disp

Página 9 - USING CUVETTES

17AluminumALUMINUMSPECIFICATIONSRange 0.00 to 1.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationL

Página 10 - INTERFERENCES

18Aluminum• Add the content of one packet of HI 93712C-0 Bleaching powder toone of the two cuvettes. Replace the cap and shake vigorously untilcomplet

Página 11 - HEALTH & SAFETY

19Ammonia MRAMMONIA MEDIUM RANGESPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.05 mg/L ±5% of reading at 25 °CTypical EMC ±0.01

Página 12 - OPERATIONAL GUIDE

2Dear Customer,Thank you for choosing a Hanna product. Please read this instruction manual carefully before using theinstrument. This manual will prov

Página 13 - CHEMICAL FORM

20Ammonia MR• Remove the cuvette.• Add 4 drops of HI 93715A-0 First Reagent (6 drops forseawater analysis). Replace the cap and mix the solution.• Add

Página 14 - SETUPSETUP

21Ammonia LRAMMONIA LOW RANGESPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/

Página 15

22Ammonia LR• Remove the cuvette.• Add 4 drops of HI 93700A-0 First Reagent (6 drops forseawater analysis). Replace the cap and mix the solution.• Add

Página 16 - HELP MODEHELP MODE

23BromineBROMINESPECIFICATIONSRange 0.00 to 8.00 mg/LResolution 0.01 mg/LAccuracy ±0.08 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLig

Página 17 - ALUMINUM

24Bromine• Remove the cuvette and add the content of one packet ofHI 93716-0 DPD reagent. Replace the cap and shake gently forabout 20 seconds to diss

Página 18

25Free ChlorineFREE CHLORINESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg/L

Página 19 - AMMONIA MEDIUM RANGE

26Free Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagen

Página 20

27Free Chlorine• Reinsert the cuvette into the instrument.• Press Read to start the reading. The instrument displays the results in mg/L of free chlo

Página 21 - AMMONIA LOW RANGE

28Total ChlorineTOTAL CHLORINESPECIFICATIONSRange 0.00 to 3.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg

Página 22

29Total Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reage

Página 23 - MEASUREMENT PROCEDURE

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment, pl

Página 24

30Total Chlorine• Press Timer and the display will show the countdown prior to the measurement or, alternatively, waitfor 2 minutes and 30 seconds and

Página 25 - FREE CHLORINE

31Chromium VI HRCHROMIUM VI HIGH RANGESPECIFICATIONSRange 0 to 1000 μg/LResolution 1 μg/LAccuracy ±5 μg/L ±4% of reading at 25 °CTypical EMC ±1 μg/LDe

Página 26 - Free Chlorine

32Chromium VI HR• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternat

Página 27

33Chromium VI LRCHROMIUM VI LOW RANGESPECIFICATIONSRange 0 to 300 μg/LResolution 1 μg/LAccuracy ±1 μg/L ±4% of reading at 25 °CTypical EMC ±1 μg/LDevi

Página 28 - TOTAL CHLORINE

34Chromium VI LR• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternat

Página 29 - Total Chlorine

35Color of WaterCOLOR OF WATERSPECIFICATIONSRange 0 to 500 PCU (Platinum Cobalt Units)Resolution 1 PCUAccuracy ±10 PCU ±5% of reading at 25 °CTypical

Página 30

36Color of Water• Fill the second cuvette up to the mark with unfiltered sampleand replace the cap. This is the apparent color.• Filter 10 mL of sampl

Página 31 - CHROMIUM VI HIGH RANGE

37Copper HRCOPPER HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/L

Página 32

38Copper HR• Remove the cuvette.• Add the content of one packet of HI 93702-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.• R

Página 33 - CHROMIUM VI LOW RANGE

39Copper LRCOPPER LOW RANGESPECIFICATIONSRange 0 to 1000 μg/LResolution 1 μg/LAccuracy ±10 μg/L ±5% of reading at 25 °CTypical EMC ±1 μg/LDeviationLig

Página 34

4PRINCIPLE OF OPERATIONAbsorption of light is a typical phenomenon of interaction between electromagnetic radiation and matter.When a light beam cross

Página 35 - COLOR OF WATER

40Copper LR• Remove the cuvette.• Add the content of one packet of HI 95747-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.• R

Página 36 - Color of Water

41IODINEIodineSPECIFICATIONSRange 0.0 to 12.5 mg/LResolution 0.1 mg/LAccuracy ±0.1 mg/L ±5% of reading at 25 °CTypical EMC ±0.1 mg/LDeviationLight Sou

Página 37 - COPPER HIGH RANGE

42Iodine• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternatively, w

Página 38

43Nickel HRNICKEL HIGH RANGESPECIFICATIONSRange 0.00 to 7.00 g/LResolution 0.01 g/LAccuracy ±0.07 ±4% of reading at 25 °CTypical EMC ±0.02 g/LDeviatio

Página 39 - COPPER LOW RANGE

44Nickel HR• Remove the cuvette and add the content of one packet ofHI 93726-0 reagent. Replace the cap and shake gentlyuntil completely dissolved.• R

Página 40

45 Nickel LRNICKEL LOW RANGESPECIFICATIONSRange 0.000 to 1.000 mg/LResolution 0.001 mg/LAccuracy ±0.010 mg/L ±7% of reading at 25 °CTypical EMC ±0.001

Página 41

46Nickel LR• Press Timer and the display will show a countdown prior to adding reagent C or, alternatively, waitfor 15 minutes. Add one packet of HI 9

Página 42

47Nickel LRINTERFERENCESInterference may be caused by:Co2+ must not be presentFe2+ must not be presentAl3+ above 32 mg/LCa2+ above 1000 mg/L (as CaCO3

Página 43 - NICKEL HIGH RANGE

48NitrateNITRATESPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±10% of reading at 25 °CTypical EMC ±0.1 mg/LDeviationLight

Página 44

49Nitrate• Replace the cap and immediately shake vigorously upand down for exactly 10 seconds. Continue to mix byinverting the cuvette gently for 50 s

Página 45 - NICKEL LOW RANGE

5Where:-log I/Io=Absorbance (A)Io=intensity of incident light beamI =intensity of light beam after absorptionελ=molar extinction coefficient at wavele

Página 46 - Nickel LR

50Nitrite HRNITRITE HIGH RANGESPECIFICATIONSRange 0 to 150 mg/LResolution 1 mg/LAccuracy ±4 mg/L ±4% of reading at 25 °CTypical EMC ±1 mg/LDeviationLi

Página 47

51Nitrite HR• Add the content of one packet of HI 93708-0 reagent.Replace the cap and shake gently until completely dissolved.• Reinsert the cuvette i

Página 48

52Nitrite LRNITRITE LOW RANGESPECIFICATIONSRange 0.00 to 1.15 mg/LResolution 0.01 mg/LAccuracy ±0.06 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/

Página 49

53Nitrite LR• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternativel

Página 50 - NITRITE HIGH RANGE

54Dissolved OxygenDISSOLVED OXYGENSPECIFICATIONSRange 0.0 to 10.0 mg/LResolution 0.1 mg/LAccuracy ±0.4 mg/L ±3% of reading at 25 °CTypical EMC ± 0.1 m

Página 51 - Nitrite HR

55Dissolved Oxygen• Let the sample stand and the flocculent agent will start to settle.• After approximately 2 minutes, when the upper half of the bot

Página 52 - NITRITE LOW RANGE

56pHpHSPECIFICATIONSRange 6.5 to 8.5 pHResolution 0.1 pHAccuracy ±0.1 pH at 25 °CTypical EMC ±0.1 pHDeviationLight Source Tungsten lamp with narrow ba

Página 53

57pH• Remove the cuvette and add 5 drops of HI 93710-0 PhenolRed Indicator. Replace the cap and mix the solution.• Reinsert the cuvette into the instr

Página 54 - DISSOLVED OXYGEN

58Phosphate HRPHOSPHATE HIGH RANGESPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±1 mg/L ±4% of reading at 25 °CTypical EMC Dev. ±0.1

Página 55

59Phosphate HR• Add the content of one packet of HI 93717B-0 Phosphate HRReagent B to the cuvette. Replace the cap and shake gentlyuntil completely di

Página 56

6INSTRUMENT DESCRIPTIONFUNCTIONAL DESCRIPTION1) Open Cuvette Lid2) Indexing mark3) Cuvette point4) Liquid Crystal Display (LCD)5) Splash proof keypad6

Página 57

60Phosphate LRPHOSPHATE LOW RANGESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±4% of reading at 25 °CTypical EMC Dev.

Página 58 - PHOSPHATE HIGH RANGE

61Phosphate LR• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternativ

Página 59

62PhosphorusPHOSPHORUSSPECIFICATIONSRange 0.0 to 15.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±4% of reading at 25 °CTypical EMC Dev. ±0.2 mg/LLight

Página 60 - PHOSPHATE LOW RANGE

63Phosphorus• Add the content of one packet of HI 93706B-0 PhosphorusReagent B (Amino Acid) to the cuvette. Replace the capand shake gently until comp

Página 61

64SilverSILVERSPECIFICATIONSRange 0.000 to 1.000 mg/LResolution 0.005 mg/LAccuracy ±0.020 mg/L ±5% of reading at 25 °CTypical EMC ± 0.001 mg/LDeviatio

Página 62 - PHOSPHORUS

65Silver• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the secondbeaker (the sample) and swirl gently to mix. Press Timer and thedisplay will s

Página 63

66Silver• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready formeasurement.• Fill a second cuvette up to the mark w

Página 64

67ZincZINCSPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight Sou

Página 65

68ZincINTERFERENCESInterference may be caused by:Aluminum above 6 mg/LCadmium above 0.5 mg/LCopper above 5 mg/LIron above 7 mg/LManganese above 5 mg/L

Página 66

69The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These messages

Página 67 - REQUIRED REAGENT

7KEYPAD DESCRIPTIONThe keypad contains 8 direct keys and 3 functional keys with the following functions:Press to perform the function displayed above

Página 68

70The analyzed data can be managed using Hanna’s product HI92000, Windows® Compatible Software.DATA MANAGEMENTWindows® is registered Trademark of &quo

Página 69 - ERRORS AND WARNINGS

71Description Range MethodAluminum 0.00 to 1.00 mg/L AluminonAmmonia MR 0.00 to 10.00 mg/L NesslerAmmonia LR 0.00 to 3.00 mg/L NesslerBromine 0.00 to

Página 70

72ACCESSORIESREAGENT SETSHI 93700-01 100 ammonia LR testsHI 93700-03 300 ammonia LR testsHI 93701-01 100 free chlorine tests (powder)HI 93701-03 300 f

Página 71 - STANDARD METHODS

73HI 93731-01 100 zinc testsHI 93731-03 300 zinc testsHI 93732-01 100 dissolved oxygen testsHI 93732-03 300 dissolved oxygen testsHI 93737-01 50 silve

Página 72 - ACCESSORIES

74WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose and

Página 73

75HANNA LITERATUREHanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. Thereference literature currently

Página 74 - WARRANTY

76Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Página 75 - HANNA LITERATURE

8The instructions listed below should be carefully followed during testing to ensure most accurate results.• Color or suspended matter in large amount

Página 76 - MAN83213 04/10

9• Proper use of the powder reagent packet:(a) use scissors to open the powder packet;(b) push the edges of the packet to form a spout;(c) pour out th