Phosphate LR
SPECIFICATIONS
Range 0.00 to 2.50 mg/L
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.01 mg/L
Light Source Tungsten lamp with narrow band interference filter @ 610 nm
Method Adaptation of the Ascorbic Acid method. The reaction between phosphate and the
reagent causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93713-0 Powder reagent 1 packet
REAGENT SETS
HI 93713-01 Reagents for 100 tests
HI 93713-03 Reagents for 300 tests
For other accessories see page 66.
MEASUREMENT PROCEDURE
Select the
Phosphate LR
method using the procedure
described in the
Method Selection
section (see page 12).
Rinse, cap and shake the cuvette several times with
unreacted sample. Fill the cuvette with 10 mL of sample
up to the mark and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cuvette and add the content of one packet of
HI 93713-0 reagent. Replace the cap and shake gently
(for about 2 minutes) until the powder is completely
dissolved.
10 mL
PHOSPHATE LOW RANGE