44
Nickel HR
Remove the cuvette and add the content of one packet of
HI 93726-0 reagent. Replace the cap and shake gently
until completely dissolved.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 1 minute and press Read. When the timer ends the meter will perform the reading.
The instrument displays concentration in g/L of nickel.
INTERFERENCES
Interference may be caused by copper.