Nickel LR
Press Timer and the display will show a countdown prior to adding reagent C or, alternatively, wait
for 15 minutes. Add one packet of HI 93740C-0 EDTA reagent to each beaker, cap and swirl to mix
until completely dissolved.
Fill one cuvette up to the mark with 10 mL of the blank.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Fill a second cuvette up to the mark with 10 mL of the reacted sample.
Insert the second cuvette into the instrument.
Press Read to start the reading. The instrument displays the results in mg/L of nickel.
Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings,
add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvette and swirl until turbidity is removed.
10 mL
10 mL