05 - ISE range
Meter status (2 chars of status byte): represents a 8 bit
hexadecimal encoding.
0x10 - temperature probe is connected
0x01 - new GLP data available
0x02 - new SETUP parameter
0x04 - out of calibration range
0x08 - the meter is in autoend point mode
Reading status (2 chars): R - in range, O - over range, U -
under range. First character corresponds to the primary
reading. Second character corresponds to mV reading.
Primary reading (corresponding to the selected range) -
11 ASCII chars, including sign and decimal point and
exponent.
Secondary reading (only when primary reading is not
mV) - 7 ASCII chars, including sign and decimal point.
Temperature reading - 7 ASCII chars, with sign and two
decimal points, always in ºC.
MDR Requests the instrument model name and firmware code (16 ASCII
chars).
GLP Requests the calibration data record.
The answer string contains:
GLP status (1 char): represents a 4 bit hexadecimal encoding.
0x01 - pH calibration available
0x02 - Rel mV calibration available
0x04 - ISE calibration available
pH calibration data (if available), which contains:
the number of calibrated buffers (1 char)
the ion charge, with sign (2 chars)
the offset, with sign and decimal point (7 chars)
the average of slopes, with sign and decimal point (7 chars)
the calibration time, yymmddhhmmss (12 chars)
buffers information (for each buffer)
type (1 char): 0 - standard, 1 - custom
status (1 char): N (new) - calibrated in last calibration;
O (old) - from an old calibration.
warnings during calibration (2 chars): 00 - no
warning, 04 - Clean Electrode warning.