Press CFM to confirm.
The second expected calibrated point is displayed.
Immerse the temperature probe into the second vessel as close as possible to
the reference thermometer. Allow a few seconds for the probe to stabilize.
Use the ARROW keys to set the calibration point value to that of the hot
water.
When the reading is stable and close to the selected calibration point,
CFM functional key is displayed.
Press CFM to confirm. The instrument returns to measurement mode.
Note: Use ARROW keys to change calibration point if necessary (±10.0 ºC)
around the point.
If the reading is not close to the selected calibration point, “Wrong”
message will blink. Change the temperature probe and restart calibra-
tion.
mV CALIBRATION
A two point calibration can be performed at 0 mV and 1800 mV.
Attach to the BNC connector a mV simulator with an accuracy of ±0.1 mV.
Enter the calibration screen. Press mV functional key.
Set 0.0 mV on the simulator.
When the reading is stable and close to the selected calibration point, the
CFM functional key is displayed.
Press CFM to confirm. The second calibration point of 1800 mV will be