It is recommended to calibrate the instrument frequently, especially if high
accuracy is required.
The ISE range should be recalibrated:
Whenever the ISE probe or ion charge is changed.
At least once a week.
After testing aggressive chemicals.
When calibration alarm time out is expired- CAL DUE tags blinks (if
feature is enabled).
Due to electrode conditioning time, the electrode must be kept immersed a few
seconds to stabilize. The user will be guided step by step during calibration
with easy to follow tags on the LCD. This will make the calibration a simple and
error-free procedure.
PROCEDURE
Select the proper ISE probe in SETUP menu (only for HI 98185) or select the
proper Ion Charge (see SETUP for details, page 33).
Note: If ISE probe is not calibrated in at least one point, the “----” will be
displayed.
Pour small quantities of the buffer solutions into clean beakers. If possible, use
plastic beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two beakers for
each standard solution. One for rinsing the electrode and one for calibration.
The instrument offers a choice of six memorized standard solutions: 0.1, 1, 10,
100, 1000, 10000 ppm and calibration up to five points. For fluoride
electrode the 2 ppm standard is also available (HI 98185 only).
Remove the protective cap from the ISE electrode.
FIVE-POINT CALIBRATION
Immerse the ISE electrode approximately 4 cm (1½”)
into the less concentrated standard solution and stir
gently.
II
II
I
SESE
SESE
SE
CALIBRATION CALIBRATION
CALIBRATION CALIBRATION
CALIBRATION
(HI 981(HI 981
(HI 981(HI 981
(HI 981
84 &84 &
84 &84 &
84 &
HI 981 HI 981
HI 981 HI 981
HI 981
8585
8585
85
))
))
)