When the reading is stable and close to the selected buffer, the CFM
functional key is displayed.
Press CFM to confirm calibration.
The calibrated value and the third expected buffer value will be dis-
played.
After the second calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½”) into the third
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected buffer, the CFM
functional key is displayed.
Press CFM to confirm calibration.
The calibrated value and the fourth expected value will be displayed.
Vista de página 17