Hanna-instruments HI 98180 Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Hanna-instruments HI 98180. Hanna Instruments HI 98180 User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - HI 98184 HI 98185

1Calibration CheckWaterproofpH/mV/ISE/TemperatureMetersHI 98180 HI 98181HI 98182 HI 98183HI 98184 HI 98185Instruction Manualwww.hannainst.com

Página 2

10HI 98182 AND HI 98183HI 98182 AND HI 98183HI 98182 AND HI 98183HI 98182 AND HI 98183HI 98182 AND HI 98183SPECIFICATIONSSPECIFICATIONSSPECIFICATIONSS

Página 3

11EGNARHp0.02ot0.2–Hp00.2ot00.2–Hp000.2ot000.2–Vm0002±5E00.1ot3-E00.1morF(.cnoc 48189IH )01E99.9ot7-E00.1morF(.cnoc 58189IH ))Fº0.842ot0.4–(Cº0.021ot0

Página 4

12OPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEINITIAL PREPARATIONThe instrument is supplied complete with rec

Página 5 - Front view

13Allow for the electrode to adjust and reading to stabilize (hourglass symbolturns off).On the pH screen are displayed:• pH reading with the selected

Página 6 - Top view

14ORP MEASUREMENTS (not for HI 98180)To perform ORP measurements, connect an optional ORP electrode (see “Accessories“section) to the instrument and t

Página 7

15Enter the ISE mode by pressing RANGE until the display changes to ISErange.Submerge the ISE electrode tip (4cm/1½”) into the sampleto be tested and

Página 8

16pppppH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONIt is recommended to calibrate the instrument frequently, especially if highac

Página 9

17• Press CAL. The instrument will display the measured pH, the LCD firstexpected buffer and the temperature reading.• If necessary, press the ARROW k

Página 10

18• When the reading is stable and close to the selected buffer, the CFMfunctional key is displayed.• Press CFM to confirm calibration.• The calibrate

Página 11

19• After the third calibration point is confirmed, immerse the pH electrodeand the temperature probe approximately 4 cm (1½”) into the fourthbuffer s

Página 12 - OPERATIONAL GUIDE

2WARRANTYWARRANTYWARRANTYWARRANTYWARRANTYDear Customer,Thank you for choosing a Hanna Instruments product.Please read this instruction manual carefull

Página 13

20• Press CFM to confirm calibration.• The instrument stores the calibration values and returns to normal mea-surement mode.FOUR, THREE or TWO-POINT C

Página 14

21ERROR SCREENSWrong bufferThe calibration cannot be confirmed.The pH reading is not close to the selected buffer. Select another buffer usingthe ARRO

Página 15

22Wrong old slopeAn inconsistency between new and previous (old) calibration is detected. Clearold calibration parameters and proceed calibration from

Página 16

23WORKING WITH CUSTOM BUFFERSIf at least one custom buffer was set in SETUP menu, it can be selected forcalibration by pressing the ARROW keys. The Cu

Página 17

24CLEAR CALIBRATIONPress Clear functional key when displayed or CLR key (HI 98180 &HI 98181) to clear old calibrations.All old calibrations, are c

Página 18

25The temperature has an effect on pH. The calibration buffer solutions areaffected by temperature changes to a lesser degree than normal solutions.Du

Página 19

26• Press CAL when the instrument is in RELATIVE mV measurement mode.The relative mV value and the temperature values are displayed.• Use the ARROW ke

Página 20

27It is recommended to calibrate the instrument frequently, especially if highaccuracy is required.The ISE range should be recalibrated:• Whenever the

Página 21

28• Press CAL. The primary LCD will displays the ion concentration in the selectedunit or “---” if not calibrated and first standard value.• If necess

Página 22

29• Press CFM to confirm calibration.• The calibrated value and the fourth expected standard value will bedisplayed.• After the third calibration poin

Página 23

3PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONThe HI 9818x instruments are state

Página 24

30Wrong slopeThe calibration cannot be confirmed.This message is displayed if slope is out of the accepted range.Slope under accepted value (30 % defa

Página 25

31GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GL

Página 26

32LAST RELATIVE mV CALIBRATION DATALast Relative mV calibration data is stored automatically after a successfulcalibration.To view the Relative mV cal

Página 27

33SETUPSETUPSETUPSETUPSETUPSetup mode allows viewing and modifying the measurement parameters.These are general SETUP parameters for all the ranges an

Página 28

34The following table lists the specific range parameters.Item Description Valid value DefaultCalibr. N

Página 29

35GENERAL PARAMETER SCREENSBacklightFocus on the Backlight item.Press Modify.Use / keys to change the intensity then press Accept to confirm.Press E

Página 30

36Auto Light OffFocus on the Auto Light Off item.Press 5, 10 or 30 to change settings.Auto Power OffFocus on the Auto Power Off item.Press Modify.Pres

Página 31

37Press Modify.Use / keys to select item.Use ARROW keys to change focused values.Press Accept to confirm new setting, or ESC to leave without changi

Página 32 - LAST ISE CALIBRATION DATA

38LanguageFocus on the Language item.Use the desired functional key to change the option. Wait until new languageis loaded. If language load fails the

Página 33 - SETUPSETUP

39Instrument IDFocus on the Instrument ID item.Press Modify.Use ARROW keys to change the instrument ID.Press Accept to confirm or ESC to exit without

Página 34 - Replace

4(–4 to 248 ºF), using HI 7662 interchangeable temperature probes ortemperature sensor inside pH electrode.These instruments can also measure with ORP

Página 35 - Contrast

40Meter informationFocus on the Meter Information item.Press Select.The meter informations are displayed:-firmware version-language version-mV and tem

Página 36 - Date/Time

41Note: If enabled “CAL DUE” warning will be displayed, the set number ofdays after calibration is over passed.First point modeFocus on the First Poin

Página 37 - Date Format

42Press Modify to set custom buffer value.Use ARROW keys to change the value.Press Accept to confirm custom buffer value or ESC to exit without saving

Página 38 - Temperature

43ISE probeFocus on the ISE probe item.Press Custom in order to set the parameters for a custom probe.Press Standard in order to select one probe from

Página 39 - Baud Rate

44Focus on Molar Weight.Press Modify in order to change molar weight.Use ARROW keys to change the value. Press Accept to confirm or ESC to exit.If Sta

Página 40 - Calibration Timeout

45Press Modify.Use ARROW key to select unit.Press Accept to confirm selection or ESC to exit.Note:• If the unit is changed or “User” is selected a war

Página 41 - Custom Buffers

46This feature allows the user to log pH, Rel mV or ISE measurements. All loggeddata can be transferred to a PC through the USB (RS232) port.The maxim

Página 42 - Out of Cal-Range Warning

47If More is pressed.Use ARROW keys to scroll between complete log information.If Delete is pressed.Use ARROW key to focus the record to be deleted an

Página 43 - Change Slope

48AAAAAutoEndutoEndutoEndutoEndutoEndTo freeze the first stable reading on the LCD press AutoEnd while the instru-ment is in measurement mode.The ”Wai

Página 44 - ISE unit

49mmmmmV V V V V andandandandand TEMPERATURE CALIBRATION TEMPERATURE CALIBRATION TEMPERATURE CALIBRATION TEMPERATURE CALIBRATION TEMPERATURE CALIBRATI

Página 45

51) Liquid Crystal Display (LCD).2) Functional keys.3) / keys to manually increase/decrease the parameters or to scroll betweenthe parameter list.4)

Página 46 - LOGGINGLOGGING

50• Press CFM to confirm.• The second expected calibrated point is displayed.• Immerse the temperature probe into the second vessel as close as possib

Página 47

51displayed.• Set 1800.0 mV on the simulator.• When the reading is stable and close to the selected calibration point, theCFM functional key is displa

Página 48

52PC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACE(((((not fornot fornot fornot fornot for HI 98180, HI 98181) HI 98180, HI 98181) HI 9818

Página 49

53OFF Is equivalent to pressing OFF keyCHR xx Change the instrument range according with the parametervalue (xx):• xx=00 pH range/0.001 resolution• xx

Página 50

54• 05 - ISE range• Meter status (2 chars of status byte): represents a 8 bithexadecimal encoding.• 0x10 - temperature probe is connected• 0x01 - new

Página 51

55• buffer value, with sign and decimal point and expo-nent (11 chars).• calibration time, yymmddhhmmss (12 chars).• electrode condition, with sign (3

Página 52

56• The ion charge (2 chars)• The ISE unit (2 chars)• The short name of the selected language (3 chars)NSLx Requests the number of logged samples (4 c

Página 53

57Notes: • “Err8” is sent if the instrument is not in measurement mode.• “Err6” is sent if the requested range is not available.• “Err4” is sent if th

Página 54

58The instrument is ready with rechargeable batteries inside.First time you start working with the instrument or when the rechargeablebatteries are ch

Página 55

59To recharge the rechargeable batteries, follow the next steps:• Connect the 12VDC power adapter to the main line and to the main line of thebattery

Página 56

614) BNC electrode connector.15) Temperature socket.Top viewTop viewTop viewTop viewTop view

Página 57

60The resistance of glass electrodes partially depends on the temperature. The lowerthe temperature, the higher the resistance. It takes more time for

Página 58

61PREPARATION PROCEDURERemove the electrode protective cap.DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normalwith electrodes and they

Página 59

62For refillable electrodes:For refillable electrodes:For refillable electrodes:For refillable electrodes:For refillable electrodes:If the filling sol

Página 60

63Follow the Storage Procedure above.pH CLEANING PROCEDURE• GeneralSoak in Hanna HI 7061 or HI 8061 GeneralCleaning Solution for approximately ½ hour.

Página 61

64TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDESMOTPMYS MELBORP NOITULOSevissecxe/esnopser

Página 62

65pH CALIBRATION SOLUTIONSHI 50004-01 pH 4.01 Buffer Solution, 20 mL sachet, 10 pcsHI 50004-02 pH 7.01 Buffer Solution, 20 mL sachet, 25 pcsHI 50007-0

Página 63 - Oil/grease

66HI 8082 3.5M KCl Electrolyte in FDA approved bottle, 4x30 mL, fordouble junction electrodesHI 8093 1M KCl+AgCl Electrolyte in FDA approved bottle, 4

Página 64 - TROUBLESHOOTING GUIDE

67HI 1131BGlass-body, single junction, refillable, combination pH electrode.Use: general purpose.HI 1330BGlass-body, semimicro, single junction, refil

Página 65 - ACCESSORIES

68HI 1332BPlastic-body (PEI), double junction, refillable, combination pH electrode.Use: general purpose.FC 100BPlastic-body (PVDF), double junction,

Página 66

69FC 911BPlastic-body (PVDF), double junction, refillable with built-in amplifier, combination pHelectrode. Use: very high humidity.HI 1413BGlass-body

Página 67

7HI 98182, HI 98183,HI 98182, HI 98183,HI 98182, HI 98183,HI 98182, HI 98183,HI 98182, HI 98183,HI 98184, AND HI 98185HI 98184, AND HI 98185HI 98184,

Página 68

70EXTENSION CABLE FOR SCREW-TYPE ELECTRODES(SCREW TO BNC ADAPTER)HI 7855/1 Extension cable 1 m (3.3') longHI 7855/3 Extension cable 3 m (9.9&ap

Página 69 - ORP ELECTRODES

71RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable for theenvironment in which they are used.Operation of th

Página 70

72SALES AND TECHNICAL SERVICE CONTACTSAustralia:Tel. (03) 9769.0666 • Fax (03) 9769.0699China:Tel. (10) 88570068 • Fax (10) 88570060Egypt:Tel. & F

Página 71

8Top viewTop viewTop viewTop viewTop view14) BNC electrode connector.15) Reference socket (HI 98184/5).16) Reference socket rubber stopper (HI 98184/5

Página 72 - MAN98185

9EGNARHp0.61ot0.2–Hp00.61ot00.2–(Vm0002± 18189IH )ylno)Fº0.842ot0.4–(Cº0.021ot0.02–NOITULOSERHp1.0Hp10.0(Vm1.0 18189IH )ylno)Fº1.0(Cº1.0YCARUCCAFº86/C

Comentários a estes Manuais

Sem comentários