Ammonia LR
SPECIFICATIONS
Range 0.0 to 10.0 mg/L
Resolution 0.01 mg/L
Accuracy ±0.1 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.1 mg/L
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92
,
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93715A-0 First Reagent 4 drops (6 drops for seawater)
HI 93715B-0 Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests
For other accessories see page 57.
MEASUREMENT PROCEDURE
ATTENTION:
Note: for sample preparation follow the IRRIGATION WATER (LR) procedure at page 20.
• Select the
Ammonia LR
method using the procedure described in the
Method Selection
section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and
replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
AMMONIA LOW RANGE
10 mL