CHLORINE, TOTAL
Chlorine, Total
10 mL
SPECIFICATIONS
Range 0.00 to 5.00 mg/L Cl
2
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L or ±4 % of reading @ 25 °C, whichever is greater
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the EPA method 330.5 and
Standard Methods for the Examination of
Water and Wastewater, 20
th
edition
, 4500-Cl G, DPD method. The reaction between
chlorine and the DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
Code Description Q.ty/test
HI 93711-0 DPD Powder Reagent 1 packet
REAGENT SETS
HI 93711-01 Reagents for 100 tests
HI 93711-03 Reagents for 300 tests
For other accessories see page 68.
MEASUREMENT PROCEDURE
Select
Chlorine, Total
method following one of the procedures described in the “Method Selection”
section (see page 15).
Take an empty vial.
Fill the vial with 10 mL of unreacted sample, then replace the cap. This is the blank.
Place the vial into the holder and push it completely down.
Press the Zero key. The instrument will perform a zero sequence and after a few seconds the display
will show “-0.0-”. Now the meter is zeroed and ready for measurement.