pH CALIBRATION WITH AUTOMATICpH CALIBRATION WITH AUTOMATIC
pH CALIBRATION WITH AUTOMATICpH CALIBRATION WITH AUTOMATIC
pH CALIBRATION WITH AUTOMATIC
TEMPERATURE COMPENSATION (HI 8614 family)TEMPERATURE COMPENSATION (HI 8614 family)
TEMPERATURE COMPENSATION (HI 8614 family)TEMPERATURE COMPENSATION (HI 8614 family)
TEMPERATURE COMPENSATION (HI 8614 family)
INITIAL PREPARATION
Connect the pH electrode to the
BNC socket.
Connect the temperature probe to
the transmitter.
Pour small quantities of pH7.01 and pH4.01 solution into two
clean beakers.
For accurate calibration use two beakers for each buffer solution, the
first one for rinsing the electrode, the second one for calibration. In
this way contamination of the buffers is minimized.
To get accurate readings, use pH7.01 and pH4.01 if you are going
to measure acid samples or pH 7.01 and pH 10.01 for alkaline
measurements.
Note: With HI 8614L the instruments
display can be used during
calibration without the need to
connect the ammeter and the
reading is directly expressed in
pH units.
The wire between the transmitter and the recorder/indicator/controller
should be a PVC insulated two wire with a wire diameter of at least
0.7 mm. This wire is fed through one of the smaller cable glands. The
maximum distance between the power supply and the amplifier is
300 m (1000'). It is not necessary to use shielded cable. The
transmitter is protected against inversion of supply voltage.
Electrode connection: connect the BNC
of the cable to the BNC socket on the
transmitter.
For HI 8614 family only: for automatic
temperature compensation, connect
the 2 terminals of the temperature probe
(HI 76608, optional) to "TEMP. PROBE"
terminals.
If automatic temperature compensation
is not required, short the "TEMP. PROBE"
terminals with a resistance according to
the external temperature:
Temperature (°C) Resistance (Ohms)
0 1634
10 1774
20 1922
30 2078
40 2242
50 2412
60 2590
A 2 kohm resistor is factory mounted for 25°C temperature compensation.