SPECIFICATIONS OF HI 8614 & HI 8614LSPECIFICATIONS OF HI 8614 & HI 8614L
SPECIFICATIONS OF HI 8614 & HI 8614LSPECIFICATIONS OF HI 8614 & HI 8614L
SPECIFICATIONS OF HI 8614 & HI 8614L
Range 4 to 20 mA / 0.00 to 14.00 pH
Resolution 0.01 mA / 0.01 pH
Accuracy
(@20°C/68°F) ±0.02 mA / ±0.02 pH
Typical EMC Deviation ±0.25 mA / ±0.2 pH
Calibration Offset: ±2.2 mA / ±2 pH
Slope: ±0.5 mA / 86 to 116%
Temperature Compensation
Fixed or automatic from 0 to 100°C
(32 to 212°F) with HI 76608 probe
Input Impedance 10
12
Ohm
Output 4-20 mA (isolated)
Installation Category I I
Protection IP 65
Environment 0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Power Supply HI 8614: 18 to 30 Vdc
HI 8614L: 20 to 36 Vdc
Load Max 500 Ohm
Dimensions 165 x 110 x 71 mm
(6.5 x 4.3 x 2.8'')
Weight 1 kg (2.2 lb.)
The transmitters use a universal BNC socket for quick and secure
connection to any electrode with a BNC connector.
For HI 8614
family models the temperature compensation is
performed by the transmitter's ATC circuitry if the temperature probe
(HI 76608, optional) is attached. If ATC is not required, it is possible
to substitute the temperature probe with a fixed resistor.
The input is isolated from the current loop to eliminate problems
related to ground loop, low insulation cables, multiple electrode
connections, and a common mode voltage of up to 100 V ensures true
differential readings.
The terminal board of the transmitter provides for connection of power
supply, pH or ORP electrodes and temperature probe (for HI 8614
family models only).
The unit is enclosed in a protective casing conforming to IP 65
standards.
Ordering codes:
Code Range LCD
HI 8614 0.00 to 14.00 pH NO
HI 8614L 0.00 to 14.00 pH YES
HI 8615 ±1000 mV NO
HI 8615L ±1000 mV YES