Hanna-instruments HI 8614N Manual do Utilizador Página 4

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 17
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 3
6 7
TERMINAL BOARD CONNECTIONSTERMINAL BOARD CONNECTIONS
TERMINAL BOARD CONNECTIONSTERMINAL BOARD CONNECTIONS
TERMINAL BOARD CONNECTIONS
Unscrew the 4 screws and remove the top cover.
There are three cable glands on the transmitter cover. The large cable gland
with the split in the rubber is for the electrode.
Connect the positive supply to the strip terminal "+VE LOOP" and the
negative supply to the terminal "-VE LOOP" of the transmitter terminal
block.
SPECIFICATIONS OF HI 8615 & HI 8615LSPECIFICATIONS OF HI 8615 & HI 8615L
SPECIFICATIONS OF HI 8615 & HI 8615LSPECIFICATIONS OF HI 8615 & HI 8615L
SPECIFICATIONS OF HI 8615 & HI 8615L
Range 4 to 20 mA / ±1000 mV
Resolution 0.01 mA / 1 mV
Accuracy
(@20°C/68°F) ±0.02 mA / ±5 mV
Typical EMC Deviation ±0.25 mA / ±15 mV
Calibration Offset: ±0.8 mA / ±100 mV
Slope: ±0.8 mA / 90 to 110%
Input Impedance 10
12
Ohm
Output 4-20 mA (isolated)
Installation Category I I
Protection IP 65
Environment 0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Power Supply HI 8615: 18 to 30 Vdc
HI 8615L: 20 to 36 Vdc
Load Max 500 Ohm
Dimensions 165 x 110 x 71 mm
(6.5 x 4.3 x 2.8'')
Weight 1 kg (2.2 lb.)
Vista de página 3
1 2 3 4 5 6 7 8 9 ... 16 17

Comentários a estes Manuais

Sem comentários