Silica
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cuvette and add the content of one packet of
HI 93705C-0 Amino acid reagent and shake until it is
completely dissolved.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement. Alternatively, wait
for exactly 3 minutes and press Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in mg/L of silica (SiO
2
).
Press orto access the second level functions.
Press the Chem Frm key to convert the result in mg/L of Silicon (Si).