Copper HR
COPPER HIGH RANGE
SPECIFICATIONS
Range 0.00 to 5.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.02 mg/L ±4% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the
EPA method.
The reaction between copper and the bicinchoninate
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93702-0 Bicinchoninate 1 packet
REAGENT SETS
HI 93702-01 Reagents for 100 tests
HI 93702-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Copper HR
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the mark) and
replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.