Calibration Alarm timeout for ISE (2 chars) - if ISE
available when x=1 (HI 3221) or when x=2 (HI 3222)
SETUP information (2 chars): 8 bit hexadecimal encoding.
0x01 - beep ON (else OFF)
0x04 - degrees Celsius (else degrees Fahrenheit)
0x08 - Offset calibration (else Point calibration)
The number of custom buffers (1 char) when x=1
The custom buffer values, with sign and decimal point,
for each defined custom buffer (7 chars) when x=1
The ID of the ISE electrode (2 chars) - if ISE available
when x=1 (HI 3221) or when x=2 (HI 3222)
The molar weight of the selected ION, with sign and
decimal point (9 ASCII characters) when x=1 (HI 3221)
or when x=2 (HI 3222)
The electrode slope (6 ASCII chars)
The ion charge (2 chars)
The ISE unit (2 chars)
The short name of the selected language (3 chars)
NSLxy Requests the number of logged samples (4 chars).
The command parameter (2 chars):
x=1 request for channel 1
x=2 request for channel 2
y=P request for pH range when x=1
y=M request for mV and Rel mV ranges
y=I request for ISE range when x=1 (HI 3221) or
when x=2 (HI 3222)
LLSxy Requests the number of lots (information about lots) on the specified
channel and range (x - channelnumber; y - range number)
xy = 11 - channel 1; range pH;
xy = 13 - channel 1; range mV;
xy = 22 - channel 2; range ISE;
xy = 23 - channel 2; range mV;
The answer string contains:
number of lots (3 chars)
ID lot (3 chars)
date (6 chars)
time (6 chars)