Hanna-instruments HI 2216 Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Hanna-instruments HI 2216. Hanna Instruments HI 2216 User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - Bench Meter

1www.hannainst.compH/mV/ISE/°CBench Meterwith Custom Buffersand Interval LoggingHI 2216Instruction Manual

Página 2

103 cm(1¼")The custom buffers allow the user to calibrate in a buffer solution differentfrom a standard one. Up to two custom buffers can be set

Página 3

11• The “ ” tag will blink on the LCD until the reading is stable.• When the reading is stable and close to the selectedbuffer, the “READY” tag will b

Página 4 - FUNCTIONAL DESCRIPTION

12• When the reading is stable and close to the selectedbuffer, the “READY” tag will be displayed and the“CFM” tag will blink.• Press CFM to confirm t

Página 5 - SPECIFICATIONS

13Clear calibration parameters and proceed with calibrationfrom the current calibration point. The instrument will keepall confirmed values during cur

Página 6 - OPERATIONAL GUIDE

14The temperature has an effect on pH. The calibration buffer solutions areaffected by temperature changes to a lesser degree than normal solutions.Du

Página 7

15It is recommended to calibrate the instrument frequently, especially ifhigh accuracy is required.The ppm range should be recalibrated:• Whenever the

Página 8

16• If necessary, press the ARROW keys to select a different standardvalue.• The “ ” tag will blink on the LCD until the readingis stable.• When the r

Página 9

17Default slope value (mV/decade):–59.16 (monovalent cation) - ion charge is 159.16 (monovalent anion) - ion charge is –1–29.58 (divalent cation) - io

Página 10 - (1¼")

18GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GLP)GOOD LABORATORY PRACTICE (GL

Página 11

19Notes: • When the instrument is not calibrated or calibration is cleared(default values loaded) there is no “expired calibration”,and the display al

Página 12

2TABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSWARRANTY ...

Página 13

20The third pH calibration buffer:The fourth pH calibration buffer:The fifth pH calibration buffer:Notes: • The “OLd” message displayed beside the pH

Página 14

21ISE CALIBRATION DATAThe ISE calibration data is stored automatically.To view ISE calibration data, press GLP when the instrument is in ISEmeasuremen

Página 15

22- or the number of days until the calibration alarm will be displayed.- or if expired (7 days ago).• The instrument ID.Notes: • If a one-point calib

Página 16

23LOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONLOGGING FUNCTIONUp to 700 LOG samples can be stored into memory.200 manually logged

Página 17

24LOT LOGGINGSelect “StAb” (stability logging) or the desired time interval.To start interval logging press LOG key while the instrument is inmeasurem

Página 18

25No ISE records:Otherwise, the instrument will display the lot number on the primary LCDline, the number of records on the secondary, “LOG” tag and “

Página 19

26• The time (hh.mm.ss) is displayed on the LCD, along with the“TIME” tag.• The date (yyyy.mm.dd) is displayed on the LCD along with the“DATE” tag.• T

Página 20

27The “dEL” message is displayed on the primary LCD and theselected record on the secondary LCD, along with ”LOG” tag.Notes: • The ARROW keys can be u

Página 21

28Setup mode allows viewing and modifying the following parameters:• Expired Calibration Alarm• Log Interval• First Custom Buffer• Second Custom Buffe

Página 22 - • The instrument ID

29LOG INTERVALPress CAL when log interval is displayed. The log interval and “CFM” willstart blinking (“MAn” for log on demand, “StAb” log on stabilit

Página 23

3PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONPRELIMINARY EXAMINATIONRemove the instrument from the pac

Página 24

30ONE-POINT CALIBRATION BEHAVIORPress CAL when “1 Pnt” message is displayed on the secondary LCD.One of the two options (“Pnt” or “OFFS”) and “CFM” wi

Página 25

31CURRENT TIMEPress CAL when the current time is displayed. The hour and “CFM” willstart blinking.Press the ARROW keys to change the hour.Press RA

Página 26

32INSTRUMENT IDPress CAL when the “InId“ is displayed. The instrument ID (“0000“ to“9999“) and “CFM” will start blinking.Press the ARROW keys to c

Página 27

33All the instruments are factory calibrated for temperature.Hanna’s temperature probes are interchangeable and no temperaturecalibration is needed wh

Página 28 - SETUPSETUP

34• Use the ARROW keys to set the reading on the secondary LCD to thatof the hot water.• When the reading is stable and close to the selectedcalibrati

Página 29

35PC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEPC INTERFACEData transmission from the instrument to the PC can be done with theHI 92000 Windows® co

Página 30

36CHR xx Change the instrument range according with the param-eter value (xx):• xx=00 pH range/0.001 resolution• xx=01 pH range/0.01 resolution• xx=02

Página 31

37• Meter status (2 chars of status byte): represents a 8bit hexadecimal encoding.• 0x10 - temperature probe is connected• 0x01 - new GLP data availab

Página 32

38• the calibration time, yymmddhhmmss (12 chars).• electrode condition, with sign (3 chars). The“-01” code means not calculated.• Rel mV calibration

Página 33 - TEMPERATURE CALIBRATION

39LODPxxx Requests the xxxth pH record logged data.LODMxxx Requests the xxxth mV/Rel mV record logged data.LODIxxx Requests the xxxth ISE record logge

Página 34

41) Liquid Crystal Display (LCD).2) CAL key, to enter and exit calibration mode.3) CFM/GLP key, to confirm calibration, different values or to display

Página 35 - PC INTERFACE

40PREPARATION PROCEDURERemove the protective cap of the pH electrode.DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normalwith electrodes. Th

Página 36

41For refillable electrodes:If the filling solution (electrolyte) is more than 2½ cm (1”) below the fillhole, add HI 7082 or HI 8082 3.5M KCl Electrol

Página 37

42CLEANING PROCEDURE• General Soak in Hanna HI 7061 or HI 8061 GeneralCleaning Solution for approximately ½ hour.• Protein Soak in Hanna HI 7073 or HI

Página 38

43TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDETROUBLESHOOTING GUIDESMOTPMYS MELBORP NOITULOSevissecxe/esnopser

Página 39

44Typical Electrode LifeAmbient Temperature 1 – 3 years90 ºC Less than 4 months120 ºC Less than 1 monthAlkaline ErrorHigh concentrations of sodium ion

Página 40

45pH BUFFER SOLUTIONSHI 70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcsHI 70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcsHI 70010P pH 10.01 Buffer Sachets,

Página 41

46ORP PRETREATMENT SOLUTIONSHI 7091L Reducing Pretreatment Solution, 500 mLHI 7092L Oxidizing Pretreatment Solution, 500 mLpH ELECTRODESAll electrodes

Página 42

47HI 1330BGlass-body, semimicro, single junction, refillable, combination pH electrode.Use: laboratory, vials.HI 1331BGlass-body, semimicro, single ju

Página 43 - TROUBLESHOOTING GUIDE

48FC 100BPlastic-body (PVDF), double junction, refillable, combination pH electrode.Use: general purpose for food industry.FC 200BPlastic-body (PVDF),

Página 44

49HI 1413BGlass-body, single junction, flat tip, Viscolene, non-refillable, combinationpH electrode. Use: surface measurement.ORP ELECTRODESHI 3131BGl

Página 45 - ACCESSORIES

5SPECIFICATIONSSPECIFICATIONSSPECIFICATIONSSPECIFICATIONSSPECIFICATIONSEGNARHp0.61ot0.2–Hp00.61ot00.2–Hp000.61ot000.2–)PRO(Vm9.999±)PRO(Vm0002±mpp0999

Página 46

50EXTENSION CABLE FOR SCREW-TYPE ELECTRODES(SCREW TO BNC ADAPTER)HI 7855/1 Extension cable 1 m (3.3') longHI 7855/3 Extension cable 3 m (9.9&ap

Página 47

51RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable for theenvironment in which they are used.Operation of th

Página 48

52Printed in EUROPE(ROMANIA)MAN2216 02/11Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Página 49 - ORP ELECTRODES

63 cm(1¼")OPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEOPERATIONAL GUIDEPOWER CONNECTIONPlug the 12 VDC adapter into the p

Página 50

73 cm(1¼")If measurements are taken successively in different samples, it is recommendedto rinse the electrode thoroughly with deionized water or

Página 51

8RELATIVE mV MEASUREMENTS• Press RANGE until “rEL” message and “mV” tag are displayed forone second. After one second the temperature will be displaye

Página 52 - MAN2216 02/11

9pppppH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONH CALIBRATIONCalibrate the instrument frequently, especially if high accuracy isrequired.The

Comentários a estes Manuais

Sem comentários