If necessary, press the ARROW keys to select a different standard
value.
The “ ” tag will blink on the LCD until the reading
is stable.
When the reading is stable and close to the selected
standard, the “READY” tag will be displayed and
the “CFM” tag will blink.
Press CFM to confirm calibration.
The calibrated value will be displayed on the primary LCD and the
second expected standard value on the secondary LCD.
Note: The instrument will automatically skip the standard used for the
first point.
After the first calibration point is confirmed, submerse the ISE electrode
approximately 3 cm (1¼”) into the second calibration solution.
If necessary, press the ARROW keys to select a different standard
value.
The “ ” tag will blink on the LCD until the reading
is stable.
When the reading is stable and close to the selected
standard, the “READY” tag will be displayed and
the “CFM” tag will blink.
Press CFM to confirm calibration.
The instrument stores the calibration value and returns to normal
measurement mode.
Notes: If the mV value is out of the mV range (±2000), “WRONG
BUF
” and “WRONG” “ ” tags will blink alternately. In this
case, check if the correct standard has been used.
If the new slope is out of the slope window, “WRONG” “
BUF
and “WRONG” “ ” tags will blink alternately. In this case,
check if the correct standard has been used.
Slope window is between ±20 mV and ±120 mV if
ion charge is not specified (undF option in SETUP menu)
or between 50% and 120% of default slope for the
corresponding ion charge.