After the second calibration point is confirmed, immerse the ISE electrode
approximately 2 cm (¾”) into the third calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the Confirm
function key is displayed.
Press Confirm to confirm calibration.
The calibrated value and the fourth expected standard value will be
displayed.
After the third calibration point is confirmed, immerse the ISE electrode
approximately 2 cm (¾”) into the fourth calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the Confirm
function key is displayed.
Press Confirm to confirm calibration.
The calibrated value and the fifth expected standard value will be dis-
played.
After the fourth calibration point is confirmed, immerse the ISE electrode
approximately 2 cm (¾”) into the fifth calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the Confirm
function key is displayed.
Press Confirm to confirm calibration. The instrument stores the calibration
value and returns to normal measurement mode.
Note: The instrument will automatically skip the standard solutions used
during calibration.