Hanna-instruments HI 3220 Manual do Utilizador Página 48

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 47
48
mV CALIBRATION
A two point calibration can be performed at 0 mV and 1800 mV.
Attach to the BNC connector a mV simulator with an accuracy of ±0.1 mV.
Enter the calibration screen. Press mV function key.
Set 0.0 mV on the simulator.
When the reading is stable and close to the selected calibration point, the
Confirm function key is displayed.
Press Confirm to confirm. The second calibration point of 1800 mV will be
displayed.
Set 1800.0 mV on the simulator.
When the reading is stable and close to the selected calibration point, the
Confirm function key is displayed.
Press Confirm to confirm. The instrument returns to calibration screen.
Press ESC to return to measurement mode.
Notes:•If the reading is not close to the selected calibration point, “WRONG”
tag will blink. Verify calibration condition or contact your vendor if
you cannot calibrate.
•Press CAL or ESC in any moment of the calibration process. The
instrument will return in the measurement mode.
Vista de página 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 67 68

Comentários a estes Manuais

Sem comentários