Note: If measurements are taken in different samples successively, it is
recommended to clean the electrode thoroughly to avoid cross-
contamination. After cleaning, rinse the electrode with water
and then with some of the solution to be tested.
MANUAL TEMPERATURE COMPENSATION
When the temperature probe is not connected, the default
temperature value is 25°C.
To temperature compensate the pH read-
ing, measure the temperature of the solution
with a ChecktempC or another accurate
thermometer. E.g. 20°C.
Press the RANGE key to enter the °C mode and adjust the
displayed value by using the arrow keys.
TAKING ORP MEASUREMENTS (for pH21 only)
The meter is factory calibrated for ORP (mV) range. Please
contact the nearest Hanna Service Center for recalibration,
if needed.
Connect an ORP electrode (optional, see "Accessories" section) to the
BNC connector on the rear panel (see page 4).
Connect the 12VDC power adapter to the instrument (see page 4)
and then to the mains.
Press the ON/OFF key to turn the meter
on.
Press the RANGE key to select the ORP
(mV) mode.
Remove the electrode protective cap, then immerse the ORP
electrode in the sample to be tested.
Note: The electrode should be submerged approximately 4 cm (1½")
into the solution.
Note: With refillable electrodes, to obtain a faster response unscrew
the refill hole cap during measurements.
Shake briefly and wait a few seconds for
the reading to stabilize, i.e. when the
hourglass symbol stops blinking. The LCD
will show the ORP (mV) value.
TAKING pH MEASUREMENTS
Make sure that the instrument has been calibrated for pH before
taking measurements (see "pH Calibration" section for details).
Connect the supplied HI 1110B pH electrode to the BNC connec-
tor on the rear panel (see page 4).
If automatic temperature compensation is required, connect the
HI 7662 temperature probe (optional) to the proper socket on
the rear panel (see page 4).
Connect the 12VDC power adapter to the instrument (see page 4)
and then to the mains.
Press the ON/OFF key to turn the meter
on.
Press the RANGE key to select the pH
mode, if necessary.
Remove the electrode protective cap, then immerse the pH
electrode and the temperature probe in the sample to be tested.
Note: If the pH electrode is dry, immerse it in HI 70300 storage
solution for at least one hour to reactivate it.
Note: The pH electrode should be submerged
approximately 4 cm (1½") into the
solution and the temperature probe (if
used) should be located as close as
possible to the electrode.
Note: With refillable electrodes, to obtain a faster response unscrew
the refill hole cap during measurements.
Shake briefly and wait a few seconds for
the reading to stabilize, i.e. when the
hourglass symbol stops blinking. The LCD
will show the pH value automatically com-
pensated for temperature variations.
After measurements, turn the meter off by pressing the ON/OFF
key, disconnect the power adapter from the mains and the pH
electrode from the instrument.
Store the electrode with a few drops of HI 70300 solution in the
protective cap.
NEVER USE DISTILLED OR DEIONIZED WATER FOR STORING
PURPOSES.