35
the calibration time, yymmddhhmmss (12 chars).
ISE calibration data (if available), which contains:
• the number of calibrated standards (1 char)
the ion charge, with sign (2 chars)
the calibration slope, with sign and decimal point (7 chars)
the calibration time, yymmddhhmmss (12 chars)
standards information (for each standard)
type (1 char): 0 - always standard solution.
status (1 char):N (new) - calibrated in last calibration;
O (old) - from an old calibration.
warnings during calibration (2 chars): 00 - no warning.
standard value, with sign and decimal point (7 chars).
calibration time, yymmddhhmmss (12 chars).
PAR Requests the setup parameters setting.
The answer string contains:
Instrument ID (4 chars)
Calibration alarm time out (2 chars)
SETUP information (2 chars): 8 bit hexadecimal encoding.
0x01 - beep ON (else OFF)
0x04 - degrees Celsius (else degrees Fahrenheit)
0x08 - Offset calibration (else Point calibration)
Auto-off/Light time (3 chars)
Auto power off time (3 chars)
The number of custom buffers (1 char)
The custom buffer values, with sign and decimal point,
for each defined custom buffer (7 chars)
The ion charge (2 chars)
NSLx Requests the number of logged samples (4 chars).
The command parameter (1 char):
P - request for pH range
M - request for mV and Rel mV ranges
I - request for ISE range
LODPxxx Requests the xxx
th
pH record logged data.
LODMxxx Requests the xxx
th
mV/Rel mV record logged data.
LODIxxx Requests the xxx
th
ISE record logged data (HI 98172).
LODPALL Requests all pH Log on demand.
LODMALL Requests all mV/Rel mV Log on demand.