Hanna-instruments HI 98170 Manual do Utilizador Página 20

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 19
20
The calibrated value will be displayed on the
primary LCD and the second expected standard
value on the secondary LCD, together with “CAL”
and “Cal Point 2” tags.
Note: The instrument will automatically skip the standard solution used for
the first point.
After the first calibration point is confirmed, immerse the ISE electrode
approximately 4 cm (1½”) into the second calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the “CFM”
tag blinks.
Press CFM to confirm calibration.
The calibrated value will be displayed on the
primary LCD and the third expected standard
value on the secondary LCD.
After the second calibration point is confirmed, immerse the ISE electrode
approximately 4 cm (1½”) into the third calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the “CFM”
tag blinks.
Press CFM to confirm calibration.
The calibrated value will be displayed on the
primary LCD and the fourth expected standard
value on the secondary LCD.
After the third calibration point is confirmed, immerse the ISE electrode
approximately 4 cm (1½”) into the fourth calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the “CFM”
tag blinks.
Press CFM to confirm calibration.
The calibrated value will be displayed on the
primary LCD and the fifth expected standard
value on the secondary LCD.
After the fourth calibration point is confirmed, immerse the ISE electrode
approximately 4 cm (1½”) into the fifth calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
LOG CFM
LOG CFM
LOG CFM
Vista de página 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 51 52

Comentários a estes Manuais

Sem comentários