Fill one graduated mixing cylinder up to the
25 mL mark with deionized water.
Add the content of one packet of HI 93746 TPTZ reagent, close
the cylinder and agitate for 30 seconds. This is the blank.
Fill a cuvet with 10 mL of the blank up
to the mark and let it sit for 3 minutes.
Replace the cap, insert the blank into
the holder and make sure that the notch
on the cap is positioned securely into
the groove.
Press ZERO and "SIP" will appear on the display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Fill one graduated mixing cylinder up to the
25 mL mark with the sample.
Add the content of one packet of HI 93746 TPTZ reagent, close
the cylinder and agitate for 30 seconds.
12
3
6
9
30"
ZERO
IRON LOW RANGE IRON LOW RANGE
IRON LOW RANGE IRON LOW RANGE
IRON LOW RANGE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
Iron LR
12
3
6
9
30"
10 mL
Vista de página 43