Hanna-instruments HI 93710 Manual do Utilizador Página 17

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 52
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 16
17
H
I9
3
7
4
8
A
B
C
Add the content of one packet of
HI 93748A ascorbic acid to each
cuvet, replace the caps and shake
gently.
Add 0.2 mL of the
HI 93748B alkaline-
cyanide reagent
solution to each cu-
vet, replace the
caps and shake gen-
tly.
Add 1 mL of the HI 93748C
0.1% PAN indicator solution
to each cuvet, replace the
caps and shake gently.
Place the cuvet with the reacted deion-
ized water (blank) into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
Wait for 2 minutes and press ZERO. "SIP" will appear during
zeroing.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Insert the second cuvet with the reacted
sample into the instrument.
ZERO
12
3
6
9
2’
HI 93742
Iron LR/Manganese LR
Vista de página 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 51 52

Comentários a estes Manuais

Sem comentários