Hanna-instruments HI 93710 Manual do Utilizador Página 40

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 52
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 39
40
Nothing is added to the third cuvet.
Place the ZERO1 sample into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
Press ZERO; "SIP" will appear during measurement, then the
instrument will display "ZER2".
Remove the ZERO1 sample and insert
the ZERO2 sample into the instru-
ment.
Press ZERO again; "SIP" will appear during measurement, then
the instrument will display the level of Magnesium hardness in
ppm CaCO
3
(together with "n").
Remove the ZERO2 sample and insert
the third cuvet into the instrument.
Press READ; "SIP" will appear during measurement, then the
instrument will display the concentration of Calcium in ppm CaCO
3
(together with "C").
Press READ again; "SIP" will appear during measurement and
the instrument will display the total hardness concentration in
ppm CaCO
3
(together with "t").
ZERO
ZERO2
ZERO
3rd CUVET
Hardness
ZERO1
Vista de página 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 51 52

Comentários a estes Manuais

Sem comentários