Hanna-instruments HI 83210-01 Manual do Utilizador Página 39

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 52
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 38
39
Silica
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Remove the cuvette and add the content of one packet
of HI 93705C-0 Amino acid reagent and shake until it
is completely dissolved.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement. Alternatively, wait for
exactly 3 minutes and press Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in mg/L of silica (SiO
2
).
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of Silicon (Si).
Vista de página 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 51 52

Comentários a estes Manuais

Sem comentários