Hanna-instruments HI 83207-01 Manual do Utilizador Página 32

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 68
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 31
32
Press Timer and the display will show the countdown prior to zeroing the blank or, alternatively, wait
for two minutes and press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cuvette.
Insert the other cuvette (# 2) with the reacted sample into the
instrument.
Press Read to start reading. The instrument displays the results in mg/L of fluoride.
Note: For wastewater or seawater samples, before performing measurements, distillation is required.
For most accurate results, use two graduated pipettes to deliver exactly 8 mL of distilled water and
8 mL of sample.
Fluoride
INTERFERENCES
Interferences may be caused by:
Alkalinity (as CaCO
3
) above 5000 mg/L
Aluminum above 0.1 mg/L
Iron, ferric above 10 mg/L
Chloride above 700 mg/L
Phosphate, ortho above 16 mg/L
Sodium hexametaphosphate above 1.0 mg/L
Sulfate above 200 mg/L
Highly colored and turbid samples may require distillation
Highly alkaline samples can be neutralized with nitric acid.
# 2
Vista de página 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 67 68

Comentários a estes Manuais

Sem comentários