Color of Water
Fill the second cuvette up to the mark with unfiltered sample
and replace the cap. This is the apparent color.
Filter 10 mL of sample through a filter with a 0.45 μm
membrane into the third cuvette, up to the 10 mL mark and
replace the cap. This is the true color.
Insert the apparent color cuvette (# 2) into the instrument
and close the lid.
Press Read to start the reading.
The meter displays the value of apparent color in PCU.
Remove the cuvette, insert the true color cuvette (# 3) into the
instrument and ensure that the notch on the cap is positioned
securely into the groove.
Press Read to start the reading. The meter displays the value of true color in PCU.