121120
Silica
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Remove the cuvette and add the content of one packet of
HI 93705C-0 Amino acid reagent and shake until it is
completely dissolved.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown
prior to the measurement. Alternatively, wait for exactly
3 minutes and press Read. When the timer ends the
meter will perform the reading. The instrument displays
concentration in mg/L of silica (SiO
2
).
Press s or t to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of Silicon (Si).
Press s or t to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Phosphate above 60 mg/L
Phosphate above 75 mg/L
Sulfide and high concentration of iron
Eliminate color and turbidity interferences by zeroing the meter with the original water sample.
Silica
SILICA
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 610 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology,
D859,
Heteropoly Blue method. The reaction between silica and reagents causes a blue tint in
the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93705A-0 Molybdate 6 drops
HI 93705B-0 Citric acid 1 packet
HI 93705C-0 Amino acid 1 packet
REAGENT SETS
HI 93705-01 Reagents for 100 tests
HI 93705-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Select the
Silica
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark).
Add 6 drops of HI 93705A-0 Molybdate reagent. Replace
the cap and swirl the solution.
Wait for 4 minutes, add the content of one packet of
HI 93705B-0 Citric acid reagent and shake until it is
completely dissolved.
Wait for 1 minute. This is the blank.
Place the cuvette into the holder and close the lid.
10 mL
Vista de página 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 67 68 69