Hanna-instruments HI 83099 Manual do Utilizador Página 27

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 69
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 26
5352
Cyanuric Acid
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 45 seconds and press Read. When the timer ends the meter will perform the reading.
The instrument displays concentration in mg/L of cyanuric acid.
Cyanuric Acid
CYANURIC ACID
SPECIFICATIONS
Range 0 to 80 mg/L
Resolution 1 mg/L
Accuracy ±1 mg/L ±15% of reading at 25 °C
Typical EMC ±1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the turbidimetric method. The reaction between cyanuric acid and the
reagent causes a white suspension in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93722-0 Powder reagent 1 packet
REAGENT SETS
HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Select the
Cyanuric Acid
method using the procedure
described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The meter will show “-0.0- when the meter is zeroed and ready for
measurement.
10 mL
Add the content of one packet of HI 93722-0 Cyanuric
Acid Reagent. Replace the cap and shake gently for
about 10 seconds (dissolution is not complete).