The Hanna HI 4009 and HI 4109 are ion selective elec-
trodes designed for the measurement of Cyanide ions in
aqueous solutions. The HI 4009 is a solid state half-cell
sensor that requires a separate reference. The HI 4109 is a
combination ion selective electrode.
IIII
IIII
II.
SpecificationsSpecifications
SpecificationsSpecifications
Specifications
Type: Solid State electrode with
a Silver halide pellet.
Ion(s) measured: Cyanide (CN
- -
- -
-
)
Measurement range: 10
-2
M to 1X 10
-6
M
260 to 0.026 ppm
Interfering ions: Silver, Sulfide and
Mercury must be absent.
Ratio of interfering ion to
CN
--
--
-
must be less than the
ratio indicated below:
1.0 for I
--
--
-
Iodide
500 for Br
--
--
-
Bromide
500 for Cl
--
--
-
Chloride
Operating Temperature: 0-80°C
Operating pH: 11 to 14 pH
(recommended)
Dimensions: 12 mm (OD) X 120 mm
nominal insertion
(0.47” X 4.72”)
Connection: BNC
WARNING:
HCN GAS IS TOXIC IF BREATHED, OR ABSORBED THROUGH
SKIN. USE OF A HOOD AND PROTECTIVE CLOTHING IS
STRONGLY ADVISED. USE ISA TO ADJUST pH ABOVE 11.