Equipment required:Equipment required:
Equipment required:Equipment required:
Equipment required:
Hanna HI 5315 Double Junction Reference Electrode
with HI 7072 Fill Solution for use with HI 4008 half-
cell.
Hanna HI 4222 pH/ISE/mV meter or other suitable
ion or pH/mV meter. (Note: log/linear graph paper is
useful if an ISE (ion) meter is not available).
Hanna HI 180 Magnetic Stirrer or equivalent with TFE
coated stirring bars (HI 731320). Note: isolate bea-
kers from stirrer motor heat by placing insulating
material such as foam or cork between them.
Hanna HI 76404 Electrode Holder or equivalent.
Plastic beakers (HI 740036P) or other suitable mea-
surement vessel.
VI. VI.
VI. VI.
VI.
Solutions RequiredSolutions Required
Solutions RequiredSolutions Required
Solutions Required
For Cupric Measurements
0.1 M cupric standard, (500 mL) HI 4008-01
ISA, (500 mL) HI 4000-00
For direct cupric measurements- Using volumetric pipettes
and glassware, make dilutions of HI 4008-01 to bracket
the concentration of the samples. Note: 0.1M cupric stan-
dard is equivalent to 6355 ppm. Standards with concen-
trations < 10
-3
M (63.6 ppm) should be prepared daily.
Store in tightly closed plastic bottle. To 100 parts standard
or sample add 2 mL of ISA.