74
Recharge the SIM card and change the expiration date or
set the “Chk. Charge” Inactive.
“
GSM No CreditGSM No Credit
GSM No CreditGSM No Credit
GSM No Credit”: appears on LCD
The number of remaining SMS is 0.
Recharge the SIM card and change the “Remaining” SMS
value, or set the “Chk.Charge” Inactive.
“
GSM Not RespondGSM Not Respond
GSM Not RespondGSM Not Respond
GSM Not Respond”: appears on LCD
The GSM modem not respond.
Check the data cable between GSM module and PCA, check
the GSM module power supply and wait for auto retry.
“
GSM Network ErrorGSM Network Error
GSM Network ErrorGSM Network Error
GSM Network Error”: appears on LCD
The GSM network not respond.
Check the GSM module antenna and wait for auto retry.
“
Pwr rstPwr rst
Pwr rstPwr rst
Pwr rst”: appears on SMS
There was a reset on the analyzer.
Check the main power supply for the cause of dropout.
“
pH calibration OldpH calibration Old
pH calibration OldpH calibration Old
pH calibration Old” on LCD and “
pH CalpH Cal
pH CalpH Cal
pH Cal” on SMS
The pH calibration is older than 1 months.
Calibrate the pH probe.
“
No pH CalibrationNo pH Calibration
No pH CalibrationNo pH Calibration
No pH Calibration” appears on LCD
The pH channel was not calibrated.
Appears after “Set Default pH Cal.” or after EEPROM
error. In the last case, call Hanna service.
“
No ORP CalibrationNo ORP Calibration
No ORP CalibrationNo ORP Calibration
No ORP Calibration” on LCD and “
No ORP CalNo ORP Cal
No ORP CalNo ORP Cal
No ORP Cal” on SMS
The ORP channel was not calibrated.
Appears after EEPROM error. Call Hanna service.
“
No TNo T
No TNo T
No T
emp. Calemp. Cal
emp. Calemp. Cal
emp. Cal” on LCD and “
No T CalNo T Cal
No T CalNo T Cal
No T Cal” on SMS
The Temperature channel was not calibrated.
Appears after EEPROM error. Call Hanna service.
ACCESORIES
ChecktempC Thermometer (-50.0 to 150.0 ºC)
pHep5 pH meter (0.00 to 14.00pH)
ORP ORP tester (+/-999 mV)
HI 1005 pH/temperature process probe
HI 2008 ORP process probe
HI 504900 GSM transmitter
HI 704731 Complete set of tubes from pressure regulator to drain PCA 310
HI 704732 Complete set of tubes from electrode holder to drain PCA 320-330
HI 70474 Peristaltic pump tube (6 pcs)
Comentários a estes Manuais