OPERATIONAL GUIDE
INITIAL PREPARATION:
The meter is supplied with 4 pieces of 1.5V AA-type alkaline batteries.
Remove the battery cover on the back of the meter and insert the
batteries while paying attention to their polarity.
Attach the fluoride electrode (#3 on
page 4), the reference electrode (#2)
and the temperature probe (#1) to their
respective connectors on top of the meter.
Press the ON/OFF key to turn the meter on.
Note: The temperature probe can also be used on its own to measure
the temperature independently.
Note: If the temperature probe is not con-
nected, the lower display will indicate
25.0°C with the “°C” symbol blinking.
MEASUREMENT OF FLUORIDE ION CONCENTRATION
Note: Please ensure the meter is calibrated prior to proceeding any
further. In order to obtain accurate results, calibrate the meter
at a temperature no more than ±3°C away from the sample
to be measured subsequently.
Note: If the fluoride electrode is new, or it has not been used for a
few days, it must be soaked in a solution of HI 70702 (10 ppm
of F
-
) for at least 3 hours prior to measurement.
Pour about 10-20 mL of the sample in a clean beaker and add the
same quantity of HI 7023 (TISAB) solution to the beaker. Mix the
solution for a few minutes.
Rinse the electrodes with deionized water and dry them attentively
with a non-abrasive cloth.
Immerse the fluoride and the reference electrode as well as the
temperature probe to a depth of at least 1 cm
(½”) in the prepared sample. Make sure that
the electrodes are not touching the bottom of
the beaker. Position the temperature probe close
to the tip of the other electrodes.
Temperature
probe
F Electrod
Ref.
11
The measured temperature is shown on the upper display.
Immerse the temperature probe in a bath at 0.0°C.
Press CAL. CAL and BUF will appear on the display with the
" " symbol blinking. The upper part of the display shows the
calibration point (0.0 °C).
When CON appears on the display it means that the value has
been recognized and the reading is stable. Press CFM to confirm.
Nota: If the reading of the instrument is too far from the expected
value (in this case 0.0°C), the " "
symbol starts blinking to advise of an
error. In this case verify if the tem-
perature of the bath is correct (for best
results, use a stirrer to agitate a solu-
tion of ice and water), and if the probe is functioning properly.
Press CAL and then press BUF to visualize the second calibration
point (50.0 °C).
Immerse the temperature probe in a bath at 50.0°C.
When CON appears it means that the value has been recognized
and the reading is stable. Press CFM to confirm.
Switch the meter off and back on again to return to normal
measurement mode.