1312
TO SCAN GLP DATA
CODE 10 - Viewing pH GLP
Select the code 10
The " pH GLP" message will scroll twice on LCD.
Electrode calibration data are verified to be within acceptable
values. If not the message "Old probe" or "Dead probe" will scroll
once on LCD, depending on calibration data.
The LCD will then display the instrument identification (ID) code.
Press to scan remaining data, in the following order:
last calibration date (DD.MM)
last calibration year
last calibration time (hh.mm)
electrode offset value in mV ("OF" appears in the lower LCD)
electrode slope in mV/pH ("SL" appears in the lower LCD)
first point calibration buffer
second point calibration buffer (only if a 2-point calibration has
been performed).
Note: Data can be viewed in reverse order pressing the key.
Press ALT and FNC to return to function selection mode.
Code 11 - Viewing mV GLP (HI98240 only)
Select code 11. The message "Volt GLP" will scroll twice across LCD.
The LCD will then display the instrument identification (ID) code.
Press to scan remaining data, in the following order:
last calibration date (DD.MM)
last calibration year
last calibration time (hh.mm)
first calibration point
second calibration point
third calibration point (if present).
Press ALT and FNC to return to setup mode.
CODE 12 - Printing GLP Data
Select code 12. The message "GLP data printing" will scroll across
LCD.
The meter will then print a complete set of the following GLP
data:
a - Instrument identification (ID) code
printed at that moment. If printout is stopped the LCD will show
the last printed sample number. It is then possible to select
another sample.
Press the ALT and FNC keys to return to setup mode.
Viewing Logged Data
Press RANGE to view data of the selected sample. Data will be
displayed in the following order:
pH or mV (mV reading available in HI98240 only)
temperature
date
time
If RANGE is pressed when the time is displayed, the LCD will pass
to the next Sample number.
It is then possible to scroll the data of the next sample by
pressing RANGE or select a different sample by the arrow keys.
to return to setup mode, press ALT and FNC when the meter
displays the sample number.
CODE 01 - Lot Summary Printout
Select the code 01.
The message "Lot table Printing" will scroll twice on LCD.
The meter will then print a complete set of information based on
the data stored in memory:
a - lot number
b - logging interval
c - starting date
d - number of samples
e - starting time
f - number and measure unit of
the logged channels.
Note: For lot 00 the number of samples only will be printed.
DELETE SET DATA
Select code 05 to delete the Log on Demand data or code 06 to
delete the Timed Log Data.
A scrolling message will be displayed.
Press the ALT and CFM keys to confirm deletion.
It is also possible to escape without data deletion pressing the ALT
and FNC keys.