VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning: Do not validate the instrument with any standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
Note: The validation is performed only for the selected parameter. For full validation of the
instrument, the following procedure must be performed for each parameter.
Turn the meter on by pressing ON/OFF.
When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
Place the CAL CHECK
TM
Standard
Cuvette A into the cuvette holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for validation.
Remove the cuvette.
Place the corresponding CAL CHECK
TM
Standard Cuvette B into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.