Hanna-instruments HI 95725C Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Hanna-instruments HI 95725C. Hanna Instruments HI 95725C User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - Cyanuric Acid / pH

1www.hannainst.comThis Instrument is inCompliance with the CE DirectivesInstruction ManualHI 95725CFree & Total ChlorineCyanuric Acid / pH ISM

Página 2 - PRELIMINARY EXAMINATION

18 19• Press ZERO/SET and “SIP” will blink onthe display.• After a few seconds the display will show“-0.0-”. Now the meter is zeroed andready for meas

Página 3 - SPECIFICATIONS

20 21• Remove the cuvet and place the CALCHECK™ Standard HI 95725-11 Cuvet Binto the holder. Ensure that the notch onthe cap is positioned securely in

Página 4 - FUNCTIONAL DESCRIPTION

22 23FACTORY CALIBRATION RESETIt is possible to restore factory calibration:DATE INPUT• Keep READ pressed to scroll to the desiredmonth number (01-12)

Página 5 - GUIDE TO DISPLAY CODES

24 25HI 95725 is warranted for two years against defects in workmanship and materials when usedfor its intended purpose and maintained according to th

Página 6

26 27Hanna publishes a wide range of catalogs and handbooks for an equally wide range ofapplications. The reference literature currently covers areas

Página 7 - OPERATIONAL GUIDE

28MAN95725CR1 04/05www.hannainst.com

Página 8 - For free chlorine measurement

2 3PRELIMINARY EXAMINATION ...3GENERAL DESCRIPTION ...

Página 9 - CYANURIC ACID

4 5The HI 95725 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a manufacturer of analytical inst

Página 10

6 7Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam crosses a substance, some

Página 11 - CALIBRATION PROCEDURE

8 9KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO: press to zero the meter prior to measurement.• READ: press for making a measuremen

Página 12

10 11The instructions listed below should be carefully followedduring testing to ensure best accuracy.• For a correct filling of the cuvet: the liquid

Página 13 - ACCESSORIES

12 13MEASUREMENTMEASUREMENTMEASUREMENTMEASUREMENTMEASUREMENTSPECIFICATIONSSPECIFICATIONSSPECIFICATIONSSPECIFICATIONSSPECIFICATIONSRange 0.00 to 5.00 m

Página 14 - HANNA LITERATURE

14 15INTERFERENCES• Positive error: - Bromine- Chlorine dioxide- Iodine- Ozone- Oxidized Manganese and Chromium• Alkalinity above 250 mg/L CaCO3 or ac

Página 15 - MAN95725CR1 04/05

16 17• Add the content of one packet ofHI 93722-0 reagent. Replace the cap andshake gently for about 10 seconds(dissolution is not complete). A milkyt