Página 1 - Ammonia Low Range ISM

1Instruction ManualHI 96700CAmmonia Low Range ISMwww.hannainst.com

Página 2 - TABLE OF CONTENTS

10Under Range: A blinking “0.00” indicates that the sampleabsorbs less light than the zero reference. Check the procedureand make sure you use the sam

Página 3 - PRELIMINARY EXAMINATION

11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c

Página 4 - ABBREVIATIONS

12MEASUREMENT PROCEDURE• Whenever the cuvette is placed into the measurement cell, itmust be dry outside, and completely free of fingerprints, oilor d

Página 5 - PRECISION AND ACCURACY

13• When the beeper sounds briefly and theLCD displays dashes, the meter is ready.The blinking “ZERO” indicates that theinstrument needs to be zeroed

Página 6 - PRINCIPLE OF OPERATION

14• Alternatively, wait for 3 minutes and 30seconds then just press READ/ /TIMER.In all cases, the lamp, cuvette and detectoricons will appear on the

Página 7 - FUNCTIONAL DESCRIPTION

15VALIDATION PROCEDUREUse the validation procedure to ensure that the instrument is properly calibrated.Warning: Do not validate the instrument with a

Página 8

16• Press CAL CHECK and the lamp, cuvette anddetector icons together with “CAL CHECK”will appear on the display, depending on themeasurement phase.• A

Página 9 - ERRORS AND WARNINGS

17CALIBRATION PROCEDURENote: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibra

Página 10

18• After a few seconds the display will show“-0.0-”. The meter is now zeroed and readyfor calibration. The blinking “READ” asks forreading calibratio

Página 11

19• Press GLP/ to edit the desired month(01-12). If the key is kept pressed, themonth number is automatically increased.• Press GLP/ to edit the des

Página 12 - MEASUREMENT PROCEDURE

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DE

Página 13

20To display the calibration date:• Press GLP/ to enter GLP mode. Thecalibration month and day will appear onthe main display and the year on theseco

Página 14

21• Press ZERO/CFM to restore the factorycalibration or press GLP/ again toabort factory calibration restore.• Press READ/ /TIMER to enter in thefact

Página 15 - VALIDATION PROCEDURE

22• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the battery cover by turning it counterclockwise.• Extract

Página 16

23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen

Página 17 - CALIBRATION PROCEDURE

24MAN96700 07/10Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Página 18

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl

Página 19

4GENERAL DESCRIPTIONABBREVIATIONSThe HI 96700 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a m

Página 20 - GLP mode

5SPECIFICATIONSPRECISION AND ACCURACYRange 0.00 to 3.00 mg/L NH3-NResolution 0.01 mg/L NH3-NAccuracy ±0.04 mg/L ±4% of reading@ 25°CTypical EMC Deviat

Página 21 - BATTERY MANAGEMENT

6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross

Página 22 - ACCESORIES

71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette ho

Página 23 - WARRANTY

810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to me

Página 24 - (ROMANIA)

9ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are alsodisplayed when the obtained values are outs