Hanna-instruments HI 96104 Manual do Utilizador Página 19

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 28
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 18
19
Alternatively, wait for 45 seconds then
just press READ/ /TIMER. In both
cases, the lamp, cuvette and detector
icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the
instrument directly displays concentration
in mg/L of cyanuric acid on the LCD.
VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning: Do not validate the instrument with any standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
Note: The validation is performed only for the selected parameter. For full validation of the
instrument, the following procedure must be performed for each parameter.
• Turn the meter on by pressing ON/OFF.
When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
Place the CAL CHECK
TM
Standard
HI 96710-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for validation.
Remove the cuvette.
Vista de página 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 27 28

Comentários a estes Manuais

Sem comentários