Hanna-instruments HI 96101 Manual do Utilizador Página 18

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 36
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 17
18
FREE CHLORINE MEASUREMENT
Fill the cuvette up to the mark with 10
mL of unreacted sample and replace the
cap.
Place the cuvette into the holder and en-
sure that the notch on the cap is positioned
securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement.
Remove the cuvette.
Add the content of one packet of
HI 93701-0 reagent. Replace the cap
and shake gently for 20 seconds (or
2 minutes in case of seawater analysis).
Press READ/ /TIMER. The lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
The instrument directly displays the
measured pH value on the Liquid Crystal
Display.
Vista de página 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 35 36

Comentários a estes Manuais

Sem comentários