SETUP
Setup mode allows viewing and modifying the instrument parameters.
The blinking “CAL” tag during setup mode suggest to press CAL/CHECK for parameters editing.
SET EPA COMPLIANCE MODE (for turbidimeter range only)
When EPA compliance reading is ON, “EPA” tag is displayed on the LCD and the reported values are
rounded to meet EPA reporting requirements.
To start edit the EPA mode, press CAL/CHECK
when “EPA” compliance reading panel is
displayed. The parameter setting and “CFM
tag will start blinking.
Press the s or t keys to set ON or OFF the
EPA compliance mode.
Press LOG/CFM to save the setting. The new
selected option of the parameter will be
displayed on the LCD. Alternatively, press CAL/
CHECK to exit without saving the new settings.
To enter/exit SETUP, press SETUP/DEL.
To select the parameter to be edit, press s or
t keys until the desired panel is displayed.
Press s or t keys also to change the value
of a parameter.
To start/stop editing a parameter, press CAL/
CHECK.
To save the new selected value of a parameter,
press LOG/CFM.