Press READ/TIMER u to start the average
reading. The display will show blinking dashes
and the icons for cuvette, detectors and lamp
will appear during measurement. When a new
intermediate value is displayed, the cuvette
icon and the measurement unit will briefly blink.
When the measurement is ended, the final
averaged result is displayed directly in NTU.
RANGE AND UNITS
HI 93414 automatically selects the correct turbidity
range to display the results with the highest
accuracy. If the measured value is higher than
1000 NTU (over range), the display will show the
maximum value blinking.
The instrument has an EPA compliance reading
mode. If this feature is activated in SETUP, “EPA
tag will appear on the LCD and the readings will be
rounded to meet EPA reporting requirements as
shown in the table.
UTN tseraeNotdroceR
0.1-0.0 50.0
01-1 1.0
04-01 1
001-04 5
004-001 01
0001-004 05
0001> 001