PASSIVE MEASURING (TESTER) MODE
This mode is intended to test an active transmitter. The HI 931002
reads the current produced by the transmitter via the 4-20 mA bus
and displays the reading in mA or pH units.
Disconnect the external power adapter from HI 931002 if present.
Turn the meter on by pressing the ON/OFF key.
Select the measurement unit by pressing the UNIT key.
Switch the selection switch to the "Measure"
mode.
Wire the calibrator "Com." and "P" connectors to the Device Under
Test (D.U.T.) input contacts (see page 4, #10 and #9 respec-
tively) using the cables provided, paying attention to the polarity
("Com." to - and "P" to + terminals).
Verify that the D.U.T. displays the same reading of the HI 931002
and calibrate the D.U.T. according to manufacturer instructions if
necessary.
IMPORTANT NOTE:
Set the selection switch to MEASURE mode only if the
D.U.T. is a transmitter and can limit current below 100mA.
WRONG OPERATION WILL BREAK THE INTERNAL FUSE.
ACTIVE DRIVE (SIMULATOR) MODE
This mode is intended to drive a chart recorder, calibrate a pressure/
current transducer or a current indicator. The HI 931002 supplies a
voltage to the 4-20 mA bus via an external connected 12-32 VDC
power supply. The calibrator regulates the current, displayed on the
LCD, as in the previous mode (Passive Drive).
Turn the meter on by pressing the ON/OFF key.
Select the measurement unit by pressing the UNIT key.
Switch the selection switch to the "Drive" mode.
Connect the external power adapter to the power
socket on the right side of the HI 931002.
Wire the calibrator "Com." and "A" connectors to the Device Under
Test (D.U.T.) input contacts (see page 4, #10 and #8 respec-
tively) using the cables provided, paying attention to the polarity
("Com." to - and "A" to + terminals).
Turn the Fine and Coarse knobs to any desired value and make
sure that the D.U.T. reflects the current value displayed on the HI
931002 LCD.
Calibrate the D.U.T. according to manufacturer instructions if necessary.
Note: the battery powers on the display only. Calibrator operational
mode is functional even without a battery or if battery is
discharged.