2-10
SETUP
Figure 6
2.3.4.3 Attaching the Beaker and Dispensing Tip
The titration reaction takes place in a closed, conical, glass beaker, sometimes called a
titration vessel, reaction vessel, titration cell or reaction cell.
The primary design features of the HI 903 titration vessel include the following:
Durability, easy to use, clean and maintain.
Conical, glass vessel body which provides strong, repeatable mixing for reaction
volumes between 50 and 150 mL.
Tightly sealing PTFE cover with low water vapor permeability and high chemical
resistivity to Karl Fischer reagents.
A sample port which can both be rapidly removed and replaced when adding solid
samples and incorporates a septum through which liquid samples can be injected.
A desiccant cartridge containing molecular sieves and/or indicating silica gel to dry
the ambient air which enters the cell as solvent, titrant and sample are added to or
removed from the vessel.
Fittings made from a highly chemically resistant material called PEEK which, in
conjuction with o-rings, create tight seals between the vessel top and the electrode,
dispensing tip, solvent and waste tubes, sample port and desiccant cartridge.
To attach the beaker assembly, see Figure 6 and follow the steps below:
Align the beaker support (D) with the base plate and attach by rotating clockwise.