9-5
METHODS OPTIMIZATION
excess iodine will result in a steep drop in electrode potential which will be interpreted by the
HI 903 as an approaching endpoint. This will in turn result in the dynamic dosing algorithm
reducing the dose size until the excess iodine has time to react. The reduced dose size
effectively interrupts the titration and adds considerable time to the titration duration. The
titration will be interrupted repeatedly in this way such that the overall titration time is
longer, even though the value of the maximum dose size is set to a large volume. The
resulting titration curve will show:
Because reaction rates are faster with two-component reagents than those observed with
one-component reagents the maximum dose volume can be set slightly higher when using
two-component systems. In the case where the maximum dose is too low, titration time will
be extended.
9.1.1.3 Timed Increment
This setting controls the amount of time between successive titrant doses.
Setting the time increment appropriately is important to ensure that the titrant has adequate
time to mix with the sample such that the electrode measures a homogeneous solution
before the titrator makes the decision on the size of the next dose of titrant.
The value of the timed increment is dependent on the type of reagent system being used.
While the default value of 1 second is compatible for use with any reagent system, titrations
using two-component reagent systems can be expedited by decreasing the time between
successive doses.
metsyStnegaeRrehcsiFlraK emuloVesoDmumixaM
smetsyStnenopmoC-enO 03ot02
µL
senotekdnasedyhedlarofsmetsyStnenopmoC-enO 52ot02
µL
enidiryphtiwdetalumrofsmetsyStnenopmoC-enO 02ot51
µL
smetsyStnenopmoC-owT 06ot04
µL
enidiryphtiwdetalumrofsmetsyStnenopmoC-owT 03ot52
µL