Hanna-instruments HI 9017 Manual do Utilizador Página 33

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 32
33
RANGE
3
PGM
mV
E.g. activation of alarm for values higher than 500 mV and lower than
-20.3 mV.
• set the instrument to the mV mode
• press 3
• press PGM.
At this stage, the left display shows "P3-1", and the right display mV
value.
• press 5
• press 0
• press 0
• press ENTER A
The left display then shows "P3-2".
• press 2
• press 0
• press decimal point
• press 3
• press +/-
• press ENTER B.
PROGRAM # 4 (HI 9017)
Setting of upper and lower temperature alarm limits.
When either limit is reached, the alarm
will be activated and the value will flash
simultaneously.
When the CLEAR key is pressed, the alarm will
be deactivated while the value will remain
flashing.
ENTER
A
500
mV
CLEAR
ENTER
B
. 3
02
+/-
mV
ºC
°C
ºC
°C