Phosphorus
PHOSPHORUS
SPECIFICATIONS
Range 0.0 to 15.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.3 mg/L ±4% of reading
Typical EMC Dev. ±0.2 mg/L
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition,
Amino Acid method. The reaction between phosphate and reagents
causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93706A-0 Molybdate 10 drops
HI 93706B-0 Amino Acid Powder 1 packet
REAGENT SETS
HI 93706-01 Reagents for 100 tests
HI 93706-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Phosphorus
method using the procedure de-
scribed in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when the
meter is zeroed and ready for measurement.
Remove the cuvette.
Add 10 drops of HI 937O6A-0 Molybdate reagent.
10 mL
x 10
Phosphate LR
Reinsert the cuvette into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and press READ.
When the timer ends the meter will perform the reading. The
instrument displays concentration in mg/L of phosphate (PO
4
3-
).
Press the or to access the second level of functions.
Press the Chem Frm functional key to convert the result in mg/L of phosphorus (P) and phosphorus
pentoxide (P
2
O
5
).
Press the or to go back to the measurement screen.
INTERFERENCES
Interference may be caused by:
Iron above 50 mg/L
Silica above 50 mg/L
Silicate above 10 mg/L
Copper above 10 mg/L
Hydrogen sulfide, arsenate, turbid sample and highly buffered samples also interfere.